摘要

目的:评估不同牙体组织厚度和窝洞深度对内冠修复中口内扫描仪(IOS)准确性的影响。

方法:使用下颌牙模模型,左侧下颌第一磨牙替换为提取的人体真实牙齿。制备具有不同深度(3.5 mm和5 mm)和剩余壁厚度(1.5 mm和0.5 mm)的内冠窝洞。使用高精度外口扫描仪扫描主模型,数据导出为STL格式。每组使用TRIOS 3 IOS进行十次口内扫描。在Geomagic Design X软件中,使用均方根(RMS)值分析真度,使用四分位距(IQR)分析精密度。使用独立样本T检验和单因素方差分析检验各组间RMS值的差异。

结果:不同窝洞深度和剩余牙体组织厚度的组别之间,RMS和IQR值存在显著差异(P< 0.001)。剩余牙体结构较薄且窝洞较深的组别表现出更大的偏差,表明扫描准确性降低。

临床意义:窝洞深度和牙体组织厚度显著影响内冠修复中口内扫描仪的准确性。

原文摘要

PURPOSE: To evaluate the effects of different tooth structure thicknesses and cavity depths on the accuracy of intraoral scanners (IOS) in endocrown restorations.

METHODS: A mandibular typodont model was used, with the left mandibular first molar replaced by an extracted real human tooth. Endocrown cavities with varying depths (3.5 mm and 5 mm) and remaining wall thicknesses (1.5 mm and 0.5 mm) were prepared. Master models were scanned using a high-precision extraoral scanner, and the data were exported in STL format. Ten intraoral scans were performed for each group using the TRIOS 3 IOS. Scanning accuracy was analyzed using the Root Mean Square (RMS) value for trueness and the Interquartile Range (IQR) for precision in Geomagic Design X software. The differences in RMS values among the groups were examined using Independent Samples T test and one-way ANOVA test.

RESULTS: Significant differences in RMS and IQR values were observed between groups with different cavity depths and remaining tooth structure thicknesses (P< 0.001). Groups with thinner remaining tooth structures and deeper cavities exhibited greater deviations, indicating reduced scanning accuracy.

CLINICAL SIGNIFICANCE: Cavity depth and tooth structure thickness significantly influenced the accuracy of intraoral scanners in endocrown restorations.

出处

American journal of dentistry 2026;39(4):211-216.